In this article, I am going to discuss the thermocouple instrument working principle, its types and applications. So let us start with our topic.
 
When two metals of different work functions are placed together, a voltage is generated at the junction. This generated voltage is proportional to the temperature and this junction is known as Thermocouple.
 

Thermocouple Instrument Working Principle

 
In a thermocouple type instrument, the electrical current to be measured is passed through a heater element. And the heat produced by the heater element is sensed by a thermocouple, as a result, an EMF is generated at its terminals which can be measured with the help of a PMMC instrument.
 

thermocouple instrument working principle

The EMF generated is proportional to the temperature and hence to the RMS value of the current passing through the heater. The scale of the PMMC instrument can be calibrated to read the current passing through the heater. This is the basic thermocouple instrument working principle.
 
These instruments can be used for both AC as well as DC applications. These instruments are very useful and accurate to measure current and voltages at very high frequencies. The thermocouple type instruments consist of the following major parts:

  • thermo-electric element and
  • indicating instrument

The EMF produced in thermocouple type instruments is DC in nature and is of the order of a few mV. Therefore, for the purpose of indication permanent magnet moving coil (PMMC) type instruments are used since they have a high degree of accuracy and sensitivity.
 
This instrument has a nearly “square law response”. Thus if we use an ordinary permanent magnet moving coil instrument having a uniform air gap, the scale will not be uniform.
 
The scale shape may be modified by suitably shaping the pole faces of the permanent magnet. To obtain a nearly uniform scale, the air gap is made non-uniform in such a way that flux density diminishes as the coil moves upscale.
 

Thermo-electric Element

 
The thermoelectric elements consist of a heater wire through which current to be measured is passed. The other part of the thermoelectric element is the thermocouple which develops an EMF proportional to the difference between the hot junction (heater in this case) and the cold reference junction.
 
The most important application of thermocouple type instruments is for the measurement of current and voltages at high frequencies, and therefore, the heater elements of such instruments should be free from skin effect.
 
The skin effect refers to a condition where at high frequencies the current is forced to pass through the outer surface of the conductor as the inductance and consequently the reactance of the inner parts of the conductor is much larger than that of the outer parts.
 
This increases the resistance of the conductor (since the current passes only through the outer part i.e. skin of the conductor and not through the inner part of the conductor whose reactance is higher than the outer part of the conductor).
 
In order to reduce the skin effect, the heater wires are designed with a small area of cross-section. In fact, in order to minimize the skin effect, a very fine wire of non-magnetic material having a very high resistivity is used for heater elements.
 

Types of Thermoelectric Elements

 
The main types of thermoelectric elements used in thermocouple instruments are as under:
 

Contact Type Thermo Electric Element

 

thermocouple type instrument working principle

The contact type thermoelectric element has a separate heater wire and the thermocouple has one junction (hot junction) in contact with the heater wire. The working of the contact type thermoelectric element is as follows:

  • Electrical energy is converted to the thermal energy in the heater element. Most of the heat energy is dissipated away and a portion of it is transferred to the hot junction.
  • The heat energy transferred to the hot junction is converted back to electrical energy by Seebeck effect.
  • The output of the thermocouple is connected to a permanent magnet moving coil instrument.
  • A portion of the electrical energy in the thermocouple circuit is converted to mechanical energy in deflecting the pointer and is stored as potential energy in the springs of the instrument.

The overall efficiency of the system is very low, of the order of a few hundredths of a percent under most favorable circumstances. This means that the power consumption of the instrument is relatively high and a sensitive DC instrument has to be used as an indicating element.
 

Non-contact Type Thermoelectric Element

 
In the non-contact type thermoelectric element, there is no electric contact between the heating element and the thermocouple, the two are separated by electrical insulation. The heat is conducted from heater to junction across the insulation. This makes the instrument sluggish and less sensitive than the contact type.
 
However, the separation between the heater and the thermocouple becomes essential for measurement of currents in circuits operated at potentials considerably above the ground potential.
 

Vacuum Type Thermoelectric Element

 
In a very low-range thermocouple instrument, the power absorption is very low. In order to attain a required output, we have to increase the efficiency and sensitivity of the thermoelectric element.
 
This is done by enclosing both the heater and thermocouple assembly in a glass tube which is then evacuated. The efficiency is increased because there is no cooling of the heater due to convection air currents.

principle of thermocouple type instrument

Figure shows a vacuum type thermo-electric element, where the heater element and the thereto-couple junction are electrically isolated from each other by a bead of insulating material. This is done to minimize the effects of capacitance between the thermocouple and the rest of the circuit at high frequencies.
 
The insulation creates a very small capacitance between the isolated circuits and is in series with the capacitance from the instrument to the ground or to other parts of the circuit. This reduces the capacitive currents which might otherwise be diverted through part of the heater element and produce errors in indication.
 
The capacitive currents may be a considerable part of the total current being measured if the instrument is used for measurement of very small currents.
 
With the advent of new materials, it is now possible to use materials for bead which have very high thermal conductivities. These instruments are very sensitive and have the same response as the contact types for currents up to 100 mA.
 
Another advantage of vacuum type is that the heater element can be operated at much higher temperatures. Vacuum type thermocouples may the used in the range 2 mA to 500 mA.  Vacuum thermocouples are suitable for current measurements at frequencies up to 100 MHz or more depending on size and resistivity of wire.
 
For voltage measurements, very great difficulties are encountered in constructing suitable series resistors for the high-frequency range. Voltmeters with resistors of usual type are used up to 10 kHz. Special multipliers for low voltages permit measurements at frequencies up to 100 MHz.
 

Bridge Type Thermoelectric Elements

 
No heater element is used in this case; instead, the current to be measured passes directly through the thermocouples and raises their temperature proportional to I2R.

thermocouple instruments

The thermocouples are connected as shown in Figure so that the resultant thermal EMFs give rise to a DC potential difference between A and B. No AC current can flow through the instrument as the bridge is balanced by keeping equal resistance for four arms.
 
This construction gives a greater output than that is given by a single thermocouple used in vacuum thereby giving a greater sensitivity and can withstand more mechanical shocks and overloads.
Bridge type elements are made with AC ratings from 100 mA to 1 A. The voltage available may be as much as 25 mV.
 

Advantages and Disadvantages of Thermocouple Type Instruments

 
Advantages
(i) The thermocouple instruments correctly indicate the RMS value of voltage or current irrespective of the waveform. The movement may be converted to a voltmeter by using multiplier resistance and a low current thermocouple. Thermo-couple voltmeters are available in ranges up to 500 V with sensitivities of 100 — 500 Ω/V.
 
(ii) These instruments are not affected by stray magnetic fields.
 
(iii) These instruments are free from frequency errors and hence they can be used over a very wide frequency range. In fact, they are primarily designed for current measurements at high frequencies.
 
In this application, they are superior to any other type of instrument both in accuracy and frequency range. Their accuracy can be as high as 1% for frequencies up to 50 MHz.  Above this frequency the effective resistance of heater wire is increased on account of skin effect, thereby reducing the accuracy.
 
Tubular designs are used for heater wire above 3 A to reduce errors on account of skin effect produced by high frequencies. For small currents up to 3 A, the heater wire is solid and very thin.
 
(iv) These instruments can measure currents in the range of 0.5 to 20 A. For higher currents, the heater element is usually situated outside the instrument case. For lower currents, a bridge type arrangement is used.
(v) These instruments have high sensitivity.
(vi) They are very useful when used as transfer instruments to calibrate DC instruments by potentiometer and standard cell.
 
Disadvantages
 
At normal rated current, the heater attains a temperature of 300°C. If we pass twice the rated current the heater would give a temperature of nearly 4 times the normal temperature, i.e., 1200°C. It is obvious that the square law rate will bring the heater to nearly its burn out temperature.
 
Thus it is very essential to protect the heater against damaging overloads. The fuses do not provide any protection as due to overload the heater wire may burn out before the fuse blows out.
 
Thus the overload capacity of a thermocouple type instrument is small as compared with other instruments and is about 150 percent of the full-scale current.
 
Thanks for reading about thermocouple instrument working principle..
 

Measuring Instruments MCQ – 3

 

#1 Which of the following instruments can be used in ac bridges for low frequencies up to 200 Hz only?

vibration galvanometer

#2 Maxwell bridge is used to measure

Inductance

#3 A bridge used for measurement of capacitance is

Schering bridge

#4 Which of the following instruments can be used in AC bridges for the frequencies up to 100 kHz?

Tunable amplifier detector

#5 Hay’s bridge can be used for measurements on

high Q inductive circuits

#6 . Capacitor type potential divider is used to extend the range of which type voltmeter?

Electrostatic voltmeter

#7 . Which of the following frequency meters can be used for the measurement of radio frequency?

Heterodyne frequency meter

#8 One wattmeter method is used to measure power in a three phase circuit when

the load is balanced in all three phases

#9 Which method can be used to measure the power in a three phase unbalanced load system?

Two wattmeter method

#10 In the two wattmeter method if power factor is 0.5, then one of the wattmeters will read

zero

#11 Highest flux density exists inside which of the following instrument?

moving coil type instrument

#12 In a moving iron instrument 12 A current causes a deflection of the needle by 60 degrees. A deflection of 15 degrees will be obtained by a current of

6 A

#13 In a moving coil instrument 12 A current causes a deflection of the needle by 60 degrees. A deflection of 15 degrees will be obtained by a current of

3 A

#14 Which of the following instruments does not use the effect of current for measurement purposes?

Electrostatic ammeter

#15 The porpoise of providing a swamping resistance in a dynamometer type moving coil instruments is

to provide equal time constant for fixed coil and moving coil, when used for AC measurements

#16 . Which of the following instruments will be suitable for the measurement of temperature of a furnace?

Optical pyrometer

#17 An electrodynamic wattmeter is not considered suitable for low power factor circuits due to

power loss in voltage coil

#18 In a megger when not in operation, the needle shows a resistance of

Infinity

finish

Results

Please share.
  •  
  •  
  •  
  •  
  •  
  •  
  •  
  •  
  •  
  •  
  •  
  •  
  •  
Please share.
  •  
  •  
  •  
  •  
  •  
  •  
  •  
  •  
  •  
  •  
  •  
  •  
  •  

Do not forget to click the “finish” button to see the correct answers and result. After clicking the “finish” button the correct answer will appear just below your wrong answer.
 

Electrical Measuring Instruments | All Posts

 

© www.yourelectricalguide.com/ thermocouple instrument working principle
Please share.
  •  
  •  
  •  
  •  
  •  
  •  
  •  
  •  
  •  
  •  
  •  
  •  
  •  

2 Comments

MFITUNDINDA VALLENCE · 09/04/2019 at

it’s my time to appreciate for the big work you have done for me as a learner your notes are easy to be understood and recalled in exam, i wish you give us more about physics 2,God bless you forever

    ajay sharma · 09/04/2019 at

    thank you

Leave a Reply

Your email address will not be published. Required fields are marked *